When evaluating the current of injury (COI), what should you do if the signal is over amplified/cropped and why?
What does the COI look like on an ECG?
localized ST elevation
What is the recommended action if unable to see any R-waves due to a large COI?
advise the surgeon to wait 5 mins as the COI will have reduced by 50% by that time
How long does it take for 50% of the COI to diminish?
about 5 min
How long does it take for the COI to recede completely?
about 10 - 15 min
What are the 2 methods for measuring R-waves at implant?
- use single R-wave/frozen assessment (COI button)