What is the purpose of the Ventricular Safety Period (VSP)?
prevents cross talk or the inhibition of ventricular pacing d/t the sensing of an atrial event
How does the Ventricular Safety Pacing (VSP) algorithm prevent crosstalk?
What is the Ventricular Safety Pacing window length?
110ms
Ventricular Safety Period (VSP) is available when device is operating in what modes?
Sensing Assurance helps ensure the appropriate detection of which rhythm(s)?
What does the Auto-Adjusting Sensitivity algorithm do?
What does Atrial Capture Management (ACM) do?
What does the Lead Monitor feature do?
measures the impedance of the connected lead(s)
What does Implant Detect do?
ensures that pacing and sensing polarities are set appropriately at the time of implant
What does the Sleep Function do?
suspends the programmed lower rate and replaces it with a sleep rate
What does the Single Chambr/Rate Hysteresis algorithm do?
How does the Auto-Adjusting Sensitivity algorithm work?
What does ACR stand for in relation to Atrial Capture Management?
Atrial Chamber Reset Method
What does AVC stand for in relation to Atrial Capture Management?
AV Conduction Method
What does Ventricular Capture Management (RA/RVCM) do?
automatically monitors ventricular pacing thresholds at periodic intervals and determines output based on programmable safety margin and minimum amplitude
What are the 2 Atrial Capture Management (ACM) methods to pacing thresholds?
Which method is used to calculate pacing thresholds if ACM senses 8 consecutive VS events and stable 1:1 AV conduction?
When will ACM use the AV Conduction method to calculate the pacing thresholds?
if ACM senses 8 consecutive VS events and stable 1:1 AV conduction
When will ACM use the Atrial Chamber Reset method to calculate the pacing thresholds?
if ACM senses 8 consecutive AS events and a HR < 87bpm
How often does Atrial Capture Management check threshold measurements?
QD at 0100 or 1 am
How often does Atrial Capture Management check threshold measurements?
QD at 0100 or 1 am
Which algorithm measures lead impedances?
lead monitor
How does RV Capture Management (RVCM) work?
How does Implant Detection work?
auto-polarity configuration delivers 3 asynchronous pulses at 85 bpm, 5 min after lead connection and again after 30 mins if first test is a failure