What is the minimum acceptable acute value for atrial sensing?
≥ 1.5 mV
What is the minimum acceptable acute value for ventricular sensing?
≥ 5.0 mV
What is the minimum acceptable acute value for atrial capture threshold?
≤ 1.5V @ 0.5ms
What is the minimum acceptable acute value for ventricular capture threshold?
≤ 1.0V @ 0.5ms